In all patients, the laser power was determined on the basis of ophthalmoscopic visibility of the treatment spot and adjusted to a spot of light-grayish color observed clinically. All procedures were performed by the same experienced clinician (M.B.). Follow-up visits were performed at day 1 and week 1 after laser treatment and at monthly intervals thereafter until month 3. Standardized ABT 199 examination procedures were repeated according to protocol at each follow-up visit. At each visit, patients underwent a complete evaluation, including standardized best-corrected
ETDRS visual acuity testing, slit-lamp examination, fundoscopy, color fundus photography, and SD-OCT
(Spectralis HRA+OCT; Heidelberg Engineering Inc, Bonn, Germany) and polarization-sensitive OCT imaging (a prototype developed at the Center for Medical Physics and Biomedical Engineering, Medical University Vienna, Austria). Fluorescein angiography was performed at baseline and at month 3. The principles of the polarization-sensitive OCT technology used in this study have been reported in detail elsewhere.17 The measurements reported in this paper were performed with an improved system that incorporates an additional scanning laser ophthalmoscope Nutlin 3a (SLO) channel for improved patient alignment.18 and 19 In Dichloromethane dehalogenase brief, the system can obtain several parameters simultaneously: intensity (as in standard OCT imaging), retardation (phase shift introduced by birefringence between 2 orthogonal linear
polarization states), and fast axis orientation (birefringent axis orientation of the sample relative to the orientation of the instrument). In addition, the spatial distribution of Stokes vectors can be measured, from which the degree of polarization uniformity (DOPU) can be derived and imaged.20 (DOPU is related to the degree of polarization known from classical optics, which can, however, not be directly measured by a coherent imaging technique such as OCT.) The instrument is operated at an A-scan rate of 20 000 A-scans per second for each polarization channel, allowing the recording of 3-dimensional data sets covering a scan field of ∼18 degrees (x) × 19 degrees (y) × 3.3 mm (z, optical distance) in 3.3 seconds. Variable raster scan patterns of 1024 × 64, 512 × 128, and 256 × 256 pixels (horizontal × vertical) can be selected. The theoretical depth resolution is ∼4 μm in tissue. The details of the segmentation algorithm used to identify the RPE were published previously.20 The algorithm is based on the intrinsic tissue properties of the RPE to scramble the polarization state of the backscattered light. This polarization scrambling causes a random variation of Stokes vectors from speckle to speckle.